      Alternate site burns can occur via current division or diversion through processes such as capacitive coupling, direct coupling, and insulation defects. Capacitive coupling is the phenomenon by which current appears to leak across insulation to a second electrically conductive structure, potentially causing inadvertent thermal injury away from the target site. These electrical discharges are uncommon with current endoscopic technology. Direct connection between the active electrode and another electrically conductive accessory can lead to serious alternate site burns, such as thermal injury to the biliary tree as a result of current conduction between a sphincterotome wire and a frayed guidewire. Endoscopic insulation failure occurs when the instrument’s shaft is compromised due to wear and tear or poor handling. Small cracks are more hazardous than easily detectable ones because they concentrate current and are more likely to cause injury. Thus, careful inspection of all accessories prior to use is mandatory.

      7.5.2 Implanted Electromagnetic Devices

      The electrical current produced by ESUs can interfere with implanted electromagnetic devices, including permanent pacemakers (PPMs), implantable cardioverter-defibrillators (ICDs), and medication pumps, resulting in undesirable effects, such as device reprogramming, pacemaker inhibition, and ICD-induced shocks.

      For PPMs, no reprogramming is needed if the patient is not dependent on the pacemaker. If the patient is pacemaker-dependent and prolonged use of electrosurgery is anticipated, the PPM should be synchronized to the asynchronous mode (VOO or DOO). For a patient with an ICD, a cardiology team or a team dedicated in the management of implantable cardiac devices should be consulted. The ICD should be reprogrammed to inactivate tachyarrhythmia detection before the use of electrosurgery. Alternatively, a magnet can be used if it can be secured over the pulse generator of the ICD. Continuous cardiopulmonary monitoring during the procedure is mandatory, with cardioversion–defibrillation equipment on standby. When feasible, the bipolar mode should be used over monopolar applications. If a monopolar output is required, the dispersive electrode (pad) should be placed away from the leads of the implanted device, with judicious use of current application. Postprocedure, the PPM or ICD should be reprogrammed to restore baseline function of the device.22,23

      7.5.3 Bowel Explosion

      Explosion has occurred in poorly prepped or unprepped colons exposed to electrosurgery. Polyethylene glycol or sodium phosphate cleansing preparations render the bowel safe for electrosurgery by decreasing the concentrations of combustible gases, mainly hydrogen, methane, and oxygen. The use of carbon dioxide for insufflation may also reduce the risk of explosion. Mannitol- or sorbitol-based preparations are contraindicated due to hydrogen gas production from these sugars by colonic bacterial fermentation. Enema preparations are insufficient for the safe performance of electrosurgery during flexible sigmoidoscopy, and explosion has been documented during unprepped proctoileoscopy and APC in a patient with subtotal colectomy.24

      7.6 Conclusion

      ESUs and electrosurgical devices are widely utilized during therapeutic endoscopy, yet remain a poorly understood technology among users, regardless of years of experience.25 A key understanding of the fundamental properties of electrosurgery aids in the educated selection of ESU settings and devices for a particular indication and desired tissue effect, and enhances the safety and performance of commonly performed procedures, such as snare polypectomy, hot biopsy, sphincterotomy, hemostasis, and APC.

      References

      [1] Tucker RD, Sievert CE, Kramolowsky EV, et al. The interaction between electrosurgical generators, endoscopic electrodes, and tissue. Gastrointest Endosc. 1992; 38(2):118–122

      [2] Morris ML. Electrosurgery in the gastroenterology suite: principles, practice, and safety. Gastroenterol Nurs. 2006; 29(2):126–132, quiz 132–134

      [3] Morris ML, Tucker RD, Baron TH, Wong Kee Song LM. Electrosurgery in gastrointestinal endoscopy: principles to practice. Am J Gastroenterol. 2009; 104(6):1563–1574

      [4] Munro MG, Abbott JA, Vilos GA, Brill AI. Radiofrequency electrical energy guidelines for authors: what’s in a name? J Minim Invasive Gynecol. 2015; 22(1):1–2

      [5] Wong Kee Song LM, Gostout CJ, Tucker RD, et al. Electrosurgery in gastrointestinal endoscopy: terminology matters. Gastrointest Endosc. 2016; 83(1):271–273

      [6] Rey JF, Beilenhoff U, Neumann CS, Dumonceau JM; European Society of Gastrointestinal Endoscopy (ESGE). European Society of Gastrointestinal Endoscopy (ESGE) guideline: the use of electrosurgical units. Endoscopy. 2010; 42(9):764–772

      [7] Tokar JL, Barth BA, Banerjee S, et al; ASGE Technology Committee. Electrosurgical generators. Gastrointest Endosc. 2013; 78(2):197–208

      [8] Singh N, Harrison M, Rex DK. A survey of colonoscopic polypectomy practices among clinical gastroenterologists. Gastrointest Endosc. 2004; 60(3):414–418

      [9] Binmoeller KF, Bohnacker S, Seifert H, et al. Endoscopic snare excision of “giant” colorectal polyps. Gastrointest Endosc. 1996; 43(3):183–188

      [10] Brooker JC, Saunders BP, Shah SG, Williams CB. Endoscopic resection of large sessile colonic polyps by specialist and non-specialist endoscopists. Br J Surg. 2002; 89(8):1020–1024

      [11] Van Gossum A, Cozzoli A, Adler M, et al. Colonoscopic snare polypectomy: analysis of 1485 resections comparing two types of current. Gastrointest Endosc. 1992; 38(4):472–475

      [12] Parra-Blanco A, Kaminaga N, Kojima T, et al. Colonoscopic polypectomy with cutting current: is it safe? Gastrointest Endosc. 2000; 51(6):676–681

      [13] Kim HS, Kim TI, Kim WH, et al. Risk factors for immediate postpolypectomy bleeding of the colon: a multicenter study. Am J Gastroenterol. 2006; 101(6):1333–1341

      [14] Burgess NG, Bahin FF, Bourke MJ. Colonic polypectomy (with videos). Gastrointest Endosc. 2015; 81(4):813–835

      [15] Bahin FF, Burgess NG, Kabir S, et al. Comparison of the histopathological effects of two electrosurgical currents in an in vivo porcine model of esophageal endoscopic mucosal resection. Endoscopy. 2016; 48(2):117–122

      [16] Fry LC, Lazenby AJ, Mikolaenko I, et al. Diagnostic quality of: polyps resected by snare polypectomy: does the type of electrosurgical current used matter? Am J Gastroenterol. 2006; 101(9):2123–2127

      [17] Veerappan SG, Ormonde D, Yusoff IF, Raftopoulos SC. Hot avulsion: a modification of an existing technique for management of nonlifting areas of a polyp (with video). Gastrointest Endosc. 2014; 80(5):884–888

      [18] Verma D, Kapadia A, Adler DG. Pure versus mixed electrosurgical current for endoscopic biliary sphincterotomy: a meta-analysis of adverse outcomes. Gastrointest Endosc. 2007; 66(2):283–290

      [19] American Society for Gastrointestinal Endoscopy Technology Committee. Mucosal ablation devices. Gastrointest Endosc. 2008; 68(6):1031–1042

      [20] Maple JT, Abu Dayyeh BK, Chauhan SS, et al; ASGE Technology Committee. Endoscopic submucosal dissection. Gastrointest Endosc. 2015; 81(6):1311–1325

      [21] Pannala R, Abu Dayyeh BK, Aslanian HR, et al; ASGE Technology Committee. Per-oral endoscopic myotomy (with video). Gastrointest Endosc. 2016; 83(6):1051–1060

      [22] Parekh PJ, Buerlein RC, Shams R, et al. An update on the management of implanted cardiac devices during electrosurgical procedures. Gastrointest Endosc. 2013; 78(6):836–841

      [23] Nelson G, Morris ML. Electrosurgery in the gastrointestinal suite: knowledge is power. Gastroenterol Nurs. 2015; 38(6):430–439

      [24] Lin OS, Biehl T, Jiranek GC, Kozarek RA. Explosion from argon cautery during proctoileoscopy of a patient with a colectomy. Clin Gastroenterol Hepatol. 2012; 10(10):1176–1178.e2